Pressure Controllers | Digital Indicators | Pressure Transducers | Dead Weight Testers | Avionics
Mensor is a manufacturer of precision pressure calibration products. They maintain their own calibration lab accredited to the ISO-17025 and ANSI Z540-1 standards by A2LA.
Mensor’s lab can provide NIST or DKD traceability with A2LA calibrations available as an option.
www.mensor.com
Pressure Controllers
|
CPC 6000 |
The Mensor CPC 6000 is a modular, dual-channel controller with up to four internal transducers. Transducer ranges from 0-10 in. H2O to 0-1500 psi are available for the CPC 6000 in gauge, bi-directional, or absolute models. Barometric reference can also be included for absolute or gauge emulation. |
|
CPC 8000 |
The Mensor CPC 8000 is a high accuracy, high pressure controller with up to two internal transducers. Barometric reference can also be included.Transducers for the CPC 8000 are available in pressure ranges from 0-2 psi to 0-23,000 psi absolute, gauge, or bi-directional. Pressure can be displayed in any number of engineering or user-defined units and barometric reference can be included for absolute or gauge emulation. |
CPC 3000 Click Here for datasheet |
The Mensor CPC 3000 is a high speed pressure controller with a single internal transducer and is designed to meet the needs of high-volume and industrial clients. Transducers for the CPC 3000 are available in pressure ranges from 0-5 psi to 0-1000 psi absolute, gauge, or bi-directional. |
Digital Indicators
CPG 2500 Click Here for datasheet |
The Mensor CPG 2500 is a precision transfer standard for calibration labs and instrument manufacturers. Using the same transducer technology available for the CPC 6000, this unit is available with 0.01% IS-50 accuracy (see IS-50 definition below). Available pressure ranges are 0-10 in. H2O to 0-10,000 psi gauge, absolute or bi-directional. |
DPG 2300 Click Here for datasheet |
The Mensor DPG 2300 is a rugged handheld calibrator used by field technicians in the process and utilities industries for process transducer calibration. Its dual-transducer configuration, min/max options and stainless steel media-isolated sensors accurate to 0.015% FS make it one of the most flexible and accurate handheld calibrators on the market today. |
DPG 2400 Click Here for datasheet |
The Mensor DPG 2400 is an economical bench-top digital pressure indicator. Accuracy is 0.03% FS and available pressure ranges are 0-10 in. H2O to 0-10,000 psi, gauge, absolute, or bi-directional. |
Pressure Transducers
CPT 6100 Click Here for datasheets |
The Mensor CPT 6100 and CPT 6180 intelligent pressure transmitters bring the high-accuracy CPC 6000 transducers to an inexpensive, compact form factor for OEM and research use. Accuracy is 0.01% FS or IS-50 (see IS-50 definition below). |
CPT 6000 Click Here for datasheet |
The Mensor CPT 6000 intelligent pressure transmitter is a compact, low-cost, high-accuracy transducer for OEM and special applications. Accuracy is 0.02% FS. |
Dead Weight Testers
CPB 3000 Click Here for datasheet |
The Mensor CPB 3000 is a compact industrial hydraulic deadweight tester suitable for calibration labs or on-site applications. Total Uncertainty 0.025% of reading Please contact us for details |
CPB 5000 Click Here for datasheet |
Mensor CPB 5000 industrial deadweight tester features a patented tool-less quick change piston-cylinder system. Total Uncertainty 0.015% or optional 0.008% of reading Please contact us for details |
CPB 5000-HP Click Here for datasheet |
Mensor CPB 5000-HP High Pressure industrial hydraulic deadweight tester with patented Quick-change Piston-Cylinder System. Total Uncertainty 0.025% or optional 0.020% of reading Please contact us for details |
Calibrator for use with CPB 5000 Click Here for datasheet |
Improves measurement accuracy of deadweight testers by compensating for the effects of Temperature, Humidity, Local Gravity, etc. Options include Piston Float Indicator and integrated voltage supply with Multimeter function for pressure transducer calibration. Please contact us for details |
Avionics
|
CPA 2501 |
The Mensor CPA 2501 is a CPG 2500 designed and programmed for manufacture, calibration and repair of Avionics and Aerospace equipment. This device is ideal for instrument manufacturers, avionics calibration labs, and wind tunnel applications. Readouts include: |
Accuracy Statements
| IS-50 |
0.01% IS-50 Accuracy means from 50% to 100% of the pressure range of a transducer, accuracy is 0.01% of Reading. |
| % Full Scale | % FS means the difference between the measured pressure and the actual pressure is within a percentage of the full range of the transducer. i.e. The applied pressure to a 0.01% FS, 1500 psi transducer reading 500 psi is 500 psi ± 0.15 psi. |
| % Reading | % RDG means the difference between the measured pressure and the actual pressure is within a percentage of the measured value.i.e. The applied pressure to a 0.01% RDG, 1500 psi transducer reading 500 psi is 500 psi ± 0.05 psi |
| Total Uncertainty vs. Accuracy | Accuracy statements take into account a defined range of errors listed by the manufacturer on the datasheet. Errors usually included in accuracy statements are: Linearity, Repeatability, Hysteresis, and sometimes Temperature Effects over a defined range. Accuracy is usually expressed as a percentage. Total Uncertainty takes into account all factors which could effect the pressure reading. Total Uncertainty is usually expressed in parts per million.0.020% is 200 ppm. |
Click Here for datasheet
Click Here for datasheet
Click Here for datasheet
Click Here for datasheet
Click Here for datasheet
Click Here for datasheets
Click Here for datasheet
Click Here for datasheet
Click Here for datasheet
Click Here for datasheet
Click Here for datasheet
Click Here for datasheet